• SwissMax 600

    SwissMax 600 – The digital Hardness Tester with integrated PC for laboratory, incoming inspection, quality assurance, manufacturing and production control, training, research and development, etc. load range of 1 ... 250 kp.

    Test methods:

    - Vickers HV nach DIN EN ISO 6507
    - Brinell HB nach DIN EN ISO 6506
    - Rockwell nach DIN EN ISO 6508
    - Kugeldruck DIN ISO 2039
    - HVT, HBT, HRN, HRT, HMR

    Some characteristics:

    - Integrated PC with Windows XP
    - 12" Monitor with Touchscreen
    - Programmable serial testing
    - Programmable test methods
    - Automatic or manual measurement
    - GMS Software upgradeable
    - Transfer data into EXCEL as desired
    - Light is provided by LED
    - The SwissMax 600 is a hardness tester for all procedures according to standards following Rockwell, Super Rockwell, Brinell, Vickers, as well as for ball and indentation hardness testing for thermo plastics. The load range goes from 1 to 250. All procedures complying with DIN EN ISO 6506, 6507, 6508, 2039, BS and ASTM are memorized.

  • Details

    The SwissMax 600...

    ...is equipped with an integrated PC operated by Windows. Test results are displayed on a 12“TFT Monitor with Touch screen, also integrated. The handling can’t be easier; the operator overlooks several windows at the same time.
    ...is operated by the GMS software, i.e. universal software for hardness testing machines, developed by GNEHM Härteprüfer AG for its new modular system to measure hardness within production lines.
    ...is equipped with a digital USB camera of 3 Mega Pixel, a LED is providing the illumination. The GMS software enables fully automatic measurements of tetragonal and circular indentation, leaving the freedom to operate these manually.
    ...monitors test loads between 9.804 N and 2451 N (1 to 250) with a precision of 0.1% over 3 million divisions, a reproducibility never reached before in this kind of measurements.
    ...erfasst Prüflasten zwischen 9.804 N und 2451 N (1 bis 250 kg) mit einer Präzision von 0.1 % über 3 Millionen Zellen.
    ...has a sensor to measure the depth (Rockwell, ball and indentation hardness testing as per ISO 2039) with a resolution of 0.1 µ. Also the resolution of the optical measurement is of 0.1 µ, even for automatic measurements.
    ...can keep record of many calibrations, as i.e., for each procedure or for particular sections of the load range and, of course, for each magnification of the optical system.
    ...indicates if it is working correctly, a control window displays any malfunction or damage.
    ...can drive or communicate with a positioning system as a cross table or a robot.
    ...provides an easy way to select programms: every test is automatically registered and archiving; it enables the operator to get back at any time to a previous procedure to pursue the series, make a single test or start a new series.
    ...keeps record of settings as light intensity and contrast; it is not necessary to optimize those again on return to the same article.
    ...can, at any time, send the data to a printer or export the data in a net.
    ...ememorizes a picture of each indentation to make it possible to control the measurements until a protocol is issued or until the data has been exported to the net.
    ...can store at any time any additional hardness test, even those that are not standardized.
    ...can be operated in any of the four languages in memory: German, French, English and Italian. Any other language is optionally available.

  • Technical Data

    Standardized procedures:

    Vickers

    DIN EN ISO 6507, ASTM E 92

    Brinell

    DIN EN ISO 6506, ASTM E 10

    Rockwell

    DIN EN ISO 6508, ASTM E 18

    Ball indentation

    DIN ISO 2039

    Load ranges:

    Vickers

    1 - 2 - 3 - 4 - 5 - 10 - 20 - 30 - 40 - 50 - 60 - 80 - 100 -120 kgf

    Brinell

    1 - 2,5 - 5 - 6,25 - 10 - 15.625 - 25 - 30 - 31.25 - 62.5 - 100 -125 - 187.5 - 250 kgf

    Rockwell

    first load 3–10kgf / main load 15 – 30 – 45 – 60 – 100 – 150 kgf

    Ball indentation

    49 - 132 - 358 - 961 N

    Procedures out of standards:

    Vickers

    Depth testing (HVT) 30 – 100 kgf

    Vickers

    Depth testing (HVT) 2,5/62,5 - 2,5/187,5 - 5/250

    Macro Rockwell

    (HMR) HMR 5/250

    - It remains possible to organize series of unconventional tests.

    Applicable indenters:

    Vickers

    Diamond pyramid 136°

    Brinell

    Ball 1 mm - 2,5 mm (5mm optional on request starting from 5kg)

    Rockwell

    Diamond cone 120°

    Ball

    1/16" - 1/8" - 1/4" - 1/2"

    Ball pressure

    Ball 5 mm°

    Magnification:

    From 35x up to 1000px with interchangeable objectives

    LED:

    Light is provided by LED.

    Dimensions:

    Test height:

    320 mm (500 mm optional)

    Overhang

    250 mm

    Weight

    150kg

    Load application:

    An electrical motor within closed automatic control loops with a load cell.

    The selection of the test load is automatically via Touch screen.

    Archiving is as well automatic.

    Results display:

    Interface

    RS 232 Standard USB

    Power supply

    230 V, 50 Hz (200 VA)

    CE - conformity:

    guaranteed

  • Scope of supply

    - Installed PC Pentium 4, Hard disk of 40 Gb and Windows

    - 12“ TFT colour monitor with Touch screen

    - Universal software for hardness testers by Gnehm (GMS).

    - LED light modulated from monitor.

    - Automatic measurement of Brinell and Vickers

    - Diamonds Rockwell, Vickers, Brinell-Indenter

    - The selection of the load, the speed of application and the dwell time are automatic as a function of the standards.

    - The setting of the dwell time can be adjusted as a function of the needs.

    Automatic correction for curved surfaces as per DIN/ISO/ASTM

    - Archiving of an undefined large number of complete procedures.

    - Automatic process followed thanks overview of the complete procedure.

    - Statistic functions are displayed immediately, with a list of digital results and a graphic to follow the trend of the successive measurements.

    - Minimum and maximum tolerances, immediate calculation of the average, the standard deviation with the Cp and Cpk coefficients.

    - Histograms

    - Conversion following DIN 50150, in scales for Rockwell A, B, C, D, E, F, G, H, R, Brinell, Vickers, Shore, Knoop and N/mm2.

    - It is always possible to specify tests that are out of standards.

    - Protocols can be sent to a printer in A4 format with data and logo of the user.

    - “ACCESS” format possible for export to other computers.

    - Digital characters can be entered from the touch screen.

    - An external USB key board and a mouse can be connected.

    - The system is open for any addition or upgrade.